77210163GB Issue 3
29
A cam operated screw in the bottom of the knife base
engages with a locating slot in the carrier. Pulling the
pivoting lever toward you releases the assembly and
enables it to be slid forward or backward on the Carrier.
The setting determines the distance between the Knife
Holder and the Specimen Head of the microtome.
3.6.1
TO fIT A KNIfE hOLDER
A LwAYS LOC K ThE h ANDwhEE L TO
PREVENT IT MOVING whEN wORKING NEAR
ThE KNIfE hOLDER
The Knife Holder is constructed of two parts - a Top and
a Base. There are three types of Top to suit the types of
knife or blade. The same Base supports all versions of Top.
To fit a Knife Holder on the carrier for disposable blades
(shown in the diagram), or a standard solid knife, first pull
the left hand pivoting lever of the Knife Holder fully toward
you.
Engage the sprung dowel at the bottom of the Knife Holder
into the circular aperture in the locating slot of the carrier.
Push the bottom of the Knife Holder away from you, then
lock in position by pushing the left hand pivoting lever away
from you.
Pull the right hand pivoting lever of the Knife Holder fully
toward you.