The rear panel must be removed to reach the transit
fixings.
To remove the rear panel, undo and remove the bolt from
the centre of the rear panel. Undo the 8 screws that secure
the rear panel, then carefully pivot the panel around its left
hand side (as seen from the rear of the instrument). Lean
the panel against the left hand side panel.
DO NOT DISTURB ThE EARTh wIRE ThAT IS
CONNECTED TO ThE REAR PANEL
Refrigeration Transit fixings:
Undo and remove the two red handles (a) from the centre
of the compressor mounting plate between the fan (b) and
the compressor (c).
Remove the two wood supports (d), coloured red, from
under the compressor mounting plate.
Make sure that the power lead from the compressor to the
junction box (e) is securely fitted.
Firmly hold the blue plastic coil of the defrost valve on the
transit plate at the back of the microtome, then cut the ties
that hold the valve (f).
With the coil of the defrost valve still connected to the white
wires, place the coil over the tube that projects upward
from the copper tube on top of the compressor cooling
matrix (g). Unpack the blue plastic components and fit the
nameplate, the washer, and then the nut to hold the coil of the
defrost valve in position. Fit the plastic plug in the hole in the
centre of the rear panel.