Thermo Scientific
Agilent 1200 Series Getting Connected Guide for LC Devices
37
6
Triggering Data Acquisition
This chapter describes how to set up the LC/MS system so that the Agilent 1200 module with
the external contact interface board triggers the Thermo Scientific mass spectrometer (MS) to
start data acquisition.
If you have not already installed the contact interface board, start with these procedures:
•
“Installing the External Contact Interface Board”
•
“Connecting the Contact Closure Cable”
To trigger data acquisition during a run
1. Choose
Start > All Programs > Thermo Xcalibur > Xcalibur
, and then click the
Instrument Setup
icon.
2. Click the icon for the LC module that contains the contact interface board (typically the
autosampler) to open its Instrument Setup window.
3. Click the
Timed Events
tab.
CAUTION
To prevent a data system failure, make sure that you disconnect the Agilent
Instant Pilot cable from the LC stack before you run data acquisitions or use the direct
control commands.
Note
Instrument methods contain the contact closure signals (to trigger data acquisition),
the chromatographic conditions, and the mass spectrometer settings for an LC/MS
application.
Note
The Instrument Setup window might take several seconds to open. If you
receive a server failure error, follow the instructions on
.
Note
If the Timed Events tab is not shown, close the Xcalibur data system and follow
the instructions in
“Setting the Configuration Options for the LC Modules”
.