Setting Up the LAN Interface
Setting the LAN Initialization Mode to “Using Default”
8
Agilent 1200 Series Getting Connected Guide for LC Devices
Thermo Scientific
4. Remove the LAN card from the module.
For instructions, refer to the manual for the Agilent G1369 LAN card. This LAN card
has a configuration switch that you can configure for the initialization modes. The card
ships with all eight switches set to the Off (down) position, which is the “Bootp”
initialization mode.
5. Set the initialization switches on the LAN card (
) as follows:
The On position for the toggle switch is up and the Off position is down. With these
switch settings, the LAN card uses the factory default settings to enable a Transmission
Control/Internet Protocol (TCP/IP) connection to the LAN interface with a Telnet
session.
Figure 7.
Configuration switches on the Agilent G1369 LAN card
6. Reinstall the LAN card.
7. Using the Category 5 Ethernet cable, connect the LAN card directly to the dedicated
Ethernet port in the data system computer.
8. Plug in the disconnected power supply cord, and then turn on the LC modules.
CAUTION
To prevent damage to an instrument, always use ESD protection when
handling electronic boards and components.
Switch #
4
5
6
Position
Off
On
On
Note
The Agilent default settings are IP address 192.168.254.11 and subnet mask
255.255.0.0.
1 2 3 4 5 6 7 8