2
Setting Up the LAN Interface
Setting the LAN Initialization Mode to “Using Default”
Thermo Scientific
Agilent 1200 Series Getting Connected Guide for LC Devices
7
Setting the LAN Initialization Mode to “Using Default”
Before you configure the LAN interface parameters, set the LAN initialization mode to use
the default IP address. The initialization mode is set with three of the eight configuration
switches located on the LAN board.
• On-board LAN interface—Use switches 6, 7, and 8 located on the back of the module
that has the LAN connector.
• Agilent G1369 LAN card—Use switches 4, 5, and 6 located on the LAN card.
Follow one of these procedures:
•
To set the on-board LAN interface’s initialization mode to use the default IP address
•
To set the LAN card’s initialization mode to use the default IP address
To set the on-board LAN interface’s initialization mode to use the default IP address
Set the initialization switches on the back panel (
) as follows:
Figure 6.
Back panel on-board LAN configuration switches set to the default IP address
To set the LAN card’s initialization mode to use the default IP address
1. For the module that contains the LAN card, turn off the power switch and unplug its
power supply cord.
2. If an Ethernet cable connects to the Ethernet port (LAN), disconnect the cable.
3. Ensure that you wear electrostatic discharge (ESD) protection.
Refer to the LC modules’ manuals for additional safety information and for information
about preventing ESD damage caused by an electrical discharge.
Switch #
6
7
8
Position
Off
On
On
CONFIG
On
position
Off
position
Switches 7 and 8 are
in the On (up) position.
CAUTION
To prevent personal injury caused by an electric shock, always turn off the
module and unplug the power supply cord from the electrical outlet before removing
the cover.