Setting Up the LAN Interface
Setting the Stored TCP/IP Address for the LAN Interface
Thermo Scientific
Agilent 1200 Series Getting Connected Guide for LC Devices
9
Setting the Stored TCP/IP Address for the LAN Interface
For an LC/MS system that includes an Agilent LC stack and a Thermo Scientific mass
spectrometer, set the IP address for the Agilent LAN interface to 172.16.0.102.
To set the stored TCP/IP address for the LAN interface
1. Follow the procedure
“Setting the LAN Initialization Mode to “Using Default””
.
2. Ensure that the LAN interface connects to the data system computer network card
labeled Surveyor MS.
3. Open the Internet Protocol Properties dialog box as follows:
a. Depending on the operating system, open the Local Area Connection Status dialog
box as follows (
):
• (Microsoft™ Windows™ 7 or later) Choose
Start > Control Panel > Network
and Internet
(or
Networking and Sharing Center
if set to the Icon view), and
then click
Local Area Connection
number
.
–or–
• (Windows XP) Choose
Start > Settings > Control Panel > Network
Connections > Local Area Connection
.
IMPORTANT
If your LC/MS setup includes only one network card, use the mass
spectrometer’s IP address.
Note
During this procedure, the mass spectrometer temporarily loses communication
with the Xcalibur data system.