ELECTRONIC PERSONAL DOSEMETER HANDBOOK
FAULTS AND RECOVERY ACTIONS
ISSUE 5
Page 76
Thermo
Counts Module
Code
Text as displayed by
EasyEPD2
EPD Action Recommended
Customer
Action
Remarks
081
Unexpected Texas chip
reset (Battery Low but
no Batt Low Status)
Reset
unit.
Return to
supplier if
recurs.
No Longer Used
082
Consecutive Drop (more
than 15 seconds)
No
Longer
Used
083
Counts not latched yet
no drop occurred
Reset into
comms
mode only
— counts
processing
disabled.
Return to
supplier
ASIC or interface to ASIC
may be faulty.
The only reason to not-latch
the counts would be that a
knock has occurred.
084
Counts Processing time
inconsistent
Reset Reset
unit.
Return to
supplier if
recurs.
Unlikely to occur unless
clock overflows. This is not
possible unless clock is
written with new value or
incremented at wrong rate.
085
Count too large (HG or
SG > 300,000) or (FB or
BC > 26,000)
Reset
- Ditto -
This fault can occur by
inputting a very large
number of counts ( only
possible under test
conditions or severe
interference).
086
Detector threshold load
failed
Reset into
comms
mode only
— counts
processing
disabled.
Reset unit.
Return to
supplier if
recurs.
Texas chip or associated
circuits may be faulty.
087
A calibration value is out
of range
Counts
processing
disabled
Return to
supplier
Display will indicate dashes,
but error code can be
extracted from error log.
An error status bit is also
set.
This fault will be picked up
on power-up, writing cal.
factors or thresholds, or
switching from OFF to ON
mode.
088
Abuse alarm (more than
15 consecutive seconds
of knocks)
None-
continues
normal
processing
Clear with
EasyEPD.
Return to
supplier if
recurs when
unit not
knocked.
Check knock detector