EPD/HB/40521/
000
ELECTRONIC PERSONAL DOSEMETER HANDBOOK
TECHNICAL DESCRIPTION
ISSUE 5
Page 44
Thermo
Additional information and warnings relating to the EPD battery are given in
Section 6.2.
4.2.3
On / Off Operating Modes
The EPD has an ON/OFF power saving facility, in which the detectors are turned
off, which is intended for short-term storage (e.g. overnight). In the power saving
mode the LCD will display OFF (i.e. the EPD is off - see also Section 3.5.5). In
the EPD OFF condition battery drain is reduced to approximately 20% of the drain
when the EPD is in the ON condition (i.e. fully operational). The EPD does not
measure radiation when it is OFF.
Alternative OFF Display
EPDs of software version 11 and later may be configured to display a 4 digit
Hexadecimal value instead of the OFF display. The 4 digit hex value may be
written using suitable software, e.g. EasyEPD2. This facility is to provide
customers with the option of having the EPD display the Calibration due date
when the EPD is Off. So for example, 2512 could mean calibrate on 25
th
December.