ELECTRONIC PERSONAL DOSEMETER HANDBOOK
EPD/HB/40521/000
FAULTS AND RECOVERY ACTIONS
ISSUE 5
Thermo
Page
67
5.4
Blank LCD Display and Other Faults
It may also be possible for the EPD to fail without producing or displaying an error
code, or causing an alarm, for example due to failure of the micro-controller or
digital circuit. A list of such faults and the recommended course of action is given
in the table below.
Symptom
Likely Cause
Recommended Action
Blank Screen
Nothing displayed on the
LCD, not even ‘Heart
Beat’ indication
Nothing displayed after
battery is inserted/
replaced.
Battery Failure
Battery cap not
corrected seated
or battery out of
specification
♦
Attempt to communicate with the
EPD via the IR communications link
to prove diagnosis.
♦
Remove old and Insert new battery
and restart unit- see section 2.1.2.
♦
Communicate with the EPD via the
IR communications link to access
stored data.
♦
Remove and replace battery cap
(see Section 2.1.2).
♦
Use a recommended battery.
Static Screen
‘Heart Beat’ indication
not flashing
1 second timer error
♦
Attempt to communicate with the
EPD via the IR communications
link.
♦
Replace battery and restart unit-
see section 2.1.2.
♦
Communicate with the EPD via the
IR communications link to access
stored data.
♦
If there is still no ‘Heart Beat’
contact supplier.
Faulty Button
Cannot mute alarm
using button
Muting Disabled
♦
Communicate with EPD via the IR
communications link.
♦
Clear alarm conditions.
♦
Enable muting if required.
No response to button
Hardware fault
♦
Communicate with the EPD via the
IR communications link to access
stored data.
♦
Contact supplier.
Faulty Sounder
No audible alarm during
Confidence Test
Sound fails to operate in
alarm condition
Hardware fault
Sounder Disabled
♦
Contact supplier.
♦
Enable sounder if required