ELECTRONIC PERSONAL DOSEMETER HANDBOOK
OPERATING INSTRUCTIONS
ISSUE 5
Page 20
Thermo
3.2.2 Dose
Alarms
Dose alarms are checked and updated every second, except at low dose rates
where this period increases to a maximum of 14 seconds. Alarms can be
adjusted or inhibited via the IR communications link.
Dose alarms are calculated against a preset alarm threshold set in the EPD.
When the dose equals or exceeds the dose threshold the LED will illuminate, the
sounder will activate and the appropriate alarm flag on the LCD will be displayed.
Note that it is dose and not total dose that is compared with the threshold.
The alarm may be muted by pressing and holding the button, unless alarm mute
(user silence) is inhibited.
The four most significant digits of the alarm thresholds can be viewed on the EPD
LCD and changed using the function button (see Section 3.5.8).
3.2.3
Dose Rate Alarms
Dose rate alarms are checked and updated every second, except at low dose
rates where this period increases to a maximum of 14 seconds. Alarms can be
adjusted or inhibited via the IR communications link.
The dose rate alarms can be made to work with hysteresis, i.e. the alarm 'off'
threshold is lower than the alarm 'on' threshold. The alarm 'off' threshold can be
set in the EPD over the same range as can the alarm 'on' threshold. As the dose
rate rises above the alarm 'on' threshold the LED will flash and the sounder is
activated (see Figure 3.3). The alarm may be muted by pressing and holding the
button, unless alarm mute (user silence) is disabled. The alarm will self-cancel
when the dose rate falls below the alarm 'off' threshold.