Thermo
Hp(0.07) Dose Display
Short-press the button to display
the current dose. This dose
parameter is used for tactical dose
monitoring (e.g. per task etc.).
Display Range: 0
µ
Sv to >16Sv
(auto ranging).
Note
: Over-range (> 1.0 Sv) is indicated with a flashing LCD display, see section
3.2.4.
Hp(0.07)/h Dose Rate Display
Short-press the button again to
display the current Hp(0.07)/h dose
rate. Only the first two significant
figures of the dose rate are
displayed, the actual reading is
rounded down
Display Range: 0
µ
Sv/h to >4.0Sv/h (auto ranging).
Note:
Over-range (> 1.0 Sv/h) is indicated with a flashing LCD display, see
section 3.2.4. This condition is latched to warn the user that the dose
value may be inaccurate.
ADS User ID Digits
An additional User ID is provided, called Approved Dosimetry Service (ADS) User
ID. This is password-protected and is used by ADS systems for “permanent”
issue EPDs. The ADS User ID may be displayed in a similar fashion to the User
ID.
Short-press the button again to display each of the remaining ADS User ID digit
displays. A non-issued EPD may have an ADS User ID = FFFFFFFFFFFF.
Typically, an ADS User ID will have fewer than 12 digits, but it must be an integer
between 1 and 999999999999. For example, an ADS User ID = 3324 will be
shown on the LCD display as follows:
(First Display) (Second Display) (Third Display)