ELECTRONIC PERSONAL DOSEMETER HANDBOOK
FAULTS AND RECOVERY ACTIONS
ISSUE 5
Page 70
Thermo
IR field until indicated ( i.e. by the system, through displays, Leds etc. ). However
the EPD communications software is tolerant of short duration loss of connection.
If the condition does occur it is normal for system software to clear the alarm and
LCD display automatically when the EPD is re-presented to the IR media. The
system software then checks the status of the EPD, completes any necessary
data write operations, advises the user what to do, and clears the alarm. The
display then returns to normal.
If the condition arises regularly then this may indicate a problem in the design of
the system software, i.e. the user interface.
If the user or system fails to clear the alarm in the way described above, then the
condition can easily be cleared using EasyEPD2.