ELECTRONIC PERSONAL DOSEMETER HANDBOOK
EPD/HB/40521/000
FAULTS AND RECOVERY ACTIONS
ISSUE 5
Thermo
Page
75
EEPROM Module
Code Text as displayed by
EasyEPD2
EPD
Action
Recommended
Customer Action
Remarks
060
EEPROM no more spare
sectors available
Write
aband
oned.
Return to supplier All 9 spare sectors have
been used. This fault is
unlikely to occur in normal
operation.
Tests have shown that the
EEPROMs are very resilient
( > 1 million bit toggles).
061
EEPROM bad sector,
requested sector number is
invalid
Return to supplier This fault is unlikely to occur.
062
EEPROM bad sector
access table entry (number
is invalid)
Return to supplier This fault is unlikely to occur,
Sector Access Table is
corrupted or EEPROM read
failed
063
EEPROM could not read
from the sector access
table area
Return to supplier Executes a Deferred Log
Fault with RECOVERED
option with the given fault
code but only if the EEPROM
fail bit is not set. If the
EEPROM fail bit is set the
fault is ignored to stop the
error store being flooded with
EEPROM faults.
064
EEPROM could not read
the next entry
Return to supplier -ditto-
065
EEPROM could not
reallocate data to a good
sector
Return to supplier -ditto-
066
EEPROM could not write
next entry
Return to supplier -ditto
Data Module
Code Text as displayed by
EasyEPD2
EPD
Action
Recommended
Customer Action
Remarks
070
Data Module Error
Reserved for future use