![Thermo Electron EPD MK2 Technical Manual Download Page 68](http://html.mh-extra.com/html/thermo-electron/epd-mk2/epd-mk2_technical-manual_1104305068.webp)
EPD/HB/40521/
000
ELECTRONIC PERSONAL DOSEMETER HANDBOOK
TECHNICAL DESCRIPTION
ISSUE 5
Page 54
Thermo
Averaging time constants used to calculate penetrating dose rate Hp(10)/h
maximise at 60 seconds below 70uSv/h and therefore the statistical errors
increase as the dose rate drops below 70uSv/h. Additional processing is used to
provide faster response but with reduced accuracy when rapid changes in dose
rate are observed. The algorithm is a compromise that reduces statistical errors
but gives a fast response time. The dose rate displays on the LCD are limited to 2
significant digits.
Longer time constants have to be used when averaging the beta dose due to the
lower sensitivity of the beta detectors. The time constant used maximises at 100
seconds below 1mSv/h. Due to limitations in the processing and memory
available, this time constant cannot be reduced when rapid changes are
observed, and the Hp(0.07)/h dose rate is approximated as the sum of Hp(10)/h
and the beta dose rate. This approximation means that at low gamma energies
(below 50 keV) the Hp(0.07)/h dose rate typically under-responds by up to 25% in
comparison to the Hp(0.07) dose response. However, above 25 keV the dose
rate response remains within
±
30% accuracy criteria.
Peak Dose Rate
The highest dose rate calculated by the EPD since this value was last cleared is
termed the peak dose rate. The peak dose rates and the times at which these
peak rates occur are recorded and can be displayed on the LCD or read over the
IR communications link.
Note
: Peak dose rate times are stored to 1 second resolution.
4.6.6 Counts
The EPD has four internal counters:
♦
HG: Hard
Gamma
♦
SG: Soft
Gamma
♦
FB: Full
Beta
♦
BC: Beta
Compensating
These record the pulses output by each of the radiation detectors.
The EPD calculates the Hp(10) and Hp(0.07) doses by weighted summing (ratios
set during calibration) of the counts received on the four detector channels.
Counts can only be read via the IR communications link, although the least
significant digits are visible on the internal counts display ( section 3.5 ).