![Thermo Electron EPD MK2 Technical Manual Download Page 41](http://html.mh-extra.com/html/thermo-electron/epd-mk2/epd-mk2_technical-manual_1104305041.webp)
ELECTRONIC PERSONAL DOSEMETER HANDBOOK
EPD/HB/40521/000
OPERATING INSTRUCTIONS
ISSUE 5
Thermo
Page
27
Note
: Over-range (> 1.0 Sv/h) is indicated with a flashing LCD display, see
section 3.2.4. This condition is latched to warn the user that the dose value
may be inaccurate.
User ID Digits
Short-press the button again to display the user identity number (User ID). This is
a 12-digit number that is shown on three separate displays - the High 4 digits, the
Middle 4 digits and the Low 4 digits. Short-press the button to display each of the
remaining User ID digit displays. The User ID cannot be changed using the
button. By convention a non-issued EPD is given a User ID = FFFFFFFFFFFF.
Typically, a User ID will have fewer than 12 digits, but it must be an integer
between 1 and 999999999999. For example, a User ID = 9012 will be shown on
the LCD display as follows:
(First Display) (Second Display) (Third Display)
In systems using 8 or fewer digits for the User ID it is usual to disable and thereby
remove the unused displays ( e.g. the first and second display in the above
example).
For software version 11 and later the first 12 characters of the wearer name may
be displayed on the EPD LCD in place of the wearer ID. The character
representation on a seven segment display is necessarily limited, see table in
Appendix C.
3.5.2
Displaying Hp(0.07) Doses,
Dose Rate and ADS User ID
To display Hp(0.07) doses, press
and hold the button until this top-
level display appears:
Three sub-displays are available (in the following sub-display order):
♦
Hp(0.07) Dose
♦
Hp(0.07)/h Dose Rate
♦
ADS User ID Digits