3.8.8 Address Setting
1. Press and hold the RMT/LCL button for approx. 3sec. The VOLT display will show the
communication port address.
2. Using the VOLT adjust encoder, check that the address can be set within the range of 0 to 30.
3.8.9 Baud Rate Setting
1. Press and hold the RMT/LCL button for approx. 3sec. The CURRENT display will show the
communication port Baud Rate.
2. Using The CURRENT adjust encoder, check that the Baud Rate can be set to 1200, 2400, 4800,
9600 and 19200.