INSTALLATION
3.1.
GENERAL
This chapter contains instructions for initial inspection, preparation for use and repackaging for
shipment. Connection to PC, setting the communication port and linking PU power supplies are
described in Chapter 7.
NOTE
PU power supplies generate magnetic fields, which might affect the
operation of other instruments. If your equipment is susceptible to
magnetic fields, do not position it adjacent to the power supply.
3.2.
PREPARATION FOR USE
In order to be operational the power supply must be connected to an appropriate AC source. The AC
source voltage should be within the power supply specification. Do not apply power before reading,
Section 3.6 and 3.7.
Table 3-1 below, describes the basic setup procedure. Follow the instructions in Table 3-1 in the
sequence given to prepare the power supply for use.
Table 3-1: Basic setup procedure
Step no.
Item
Description
Reference
1
Inspection
Initial physical inspection of the power supply
Section 3.3
2
Installation
Installing the power supply,
Ensuring adequate ventilation.
Section 3.4
Section 3.5
3
AC source
AC source requirements
Connecting the power supply to the AC source.
Section 3.6
Section 3.7
4
Test
Turn-on checkout procedure.
Section 3.8
5
Load connection
Wire size selection. Local /Remote sensing.
Single or multiple loads.
Section 3.9
6
Default setting
The power supply setting at shipment.
Section 7.2.1
3.3.
INITIAL INSPECTION
Prior to shipment this power supply was inspected and found free of mechanical or electrical defects.
Upon unpacking of the power supply, inspect for any damage, which may have occured in transit.
The inspection should confirm that there is no exterior damage to the power supply such as broken
knobs or connectors and that the front panel and meters face are not scratched or cracked. Keep all
packing material until the inspection has been completed. If damage is detected, file a claim with
carrier immediately and notify the TEXIO agent nearest you.
3.4.
RACK MOUNTING
Contact the TEXIO When Rack Mounting Kit is required.
3.5.
LOCATION, MOUNTING AND COOLING
This power supply is fan cooled. The air intake is at the front panel and the exhaust is at the rear
panel. Upon installation allow cooling air to reach the front panel ventilation inlets. Allow minimum
15cm of unrestricted air space at the front and the rear of the unit.
The power supply should be used in an area that the ambient temperature does not 50℃.