5. OPERATION PROCEDURES
5-1
.
Independent Operation
*
Use the switches and terminals on the front panel to use the PA-B
series alone.
*
Make sure that the switches on the front panel are set as shown
below.
OUTPUT switch
:
F (front) position
MASTER/SLAVE switch
:
M (master) position
Various control ON/OFF switches
:
OFF positions
5-2
.
Remote Sensing
*
When the PA-B series is connected with a load, the voltage at the
load terminals is subject to load variance, which depends on the
voltage drop caused by contact resistance of the output terminals or
resistance of the load cable conductors.
Set the remote sensing selector switch on the front panel to the
REM position, connect wires to the (+S) and (-S) remote sensing
terminals and carry out remote sensing operation, if it is necessary
to set the voltage at the load terminals finely.
*
The PA-B series compensates for voltage drop of the load cable up
to 0.5 V (one way).
*
Do not set the selector switch to the REM position in normal
operation without using the remote sensing function.