20
A. Voltage setting table
WARNING
NOTE: Be sure to disconnect the AC power cable before setting the
input AC voltage.
Be sure to use the proper fuse as shown below.
10-5
18-1.2
18-2
18-3
18-5
36-1.2
36-2
36-3
Fuse
(time-lag)
100/200 VAC
3A/
2A
1.5A/
0.8A
2A/
1.25A
3A/
2A
5A/
3A
2.5A/
1.5A
3A/
2A
5A/
3A
80-1
120-0.6 160-0.4 250-0.25 250-0.42
350-0.2
600-0.1
Fuse
(time-lag)
100/200 VAC
3A/2A
3A/2A
3A/2A
3A/2A
4A/2.5A
3A/2A
3A/
2A
26. F
use holder
The AC input fuse is built in this holder.
27. Frame GND terminal screw
Use this screw to ground the frame of the PA-B series. Never
use any other screw. Using any other screw may cause
short-circuit inside the PA-B series, resulting in malfunctioning or
breakdown.