Using the USB Interface Connection
7
Using the USB Interface Connection
The on-board MSP430F5529 USB microcontroller (U4) provides an I
2
C host interface to the LMK61E2
slave device. The device registers can be controlled via USB using the GUI platform running on a Host
PC.
8
EVM Quick Start Guide
The following guide allows the user to quickly configure the LMK61E2 to evaluate performance and device
flexibility
•
1. Set the Control Pin jumpers to the default configuration as noted in Table 1Table 3.
•
2. Confirm the EVM Default power configuration is set per Table 2 to power the LMK61E2 in USB
POWER mode using the on-board LDO regulators and power supplied over USB.
•
3. Observe any active output clock on OUTN and OUTP SMA ports.
–
3.1. Default EVM configuration is AC coupled LVPECL as noted in Table 4.
–
3.2. Use 50-ohm coax cables to connect the test equipment to the output SMA ports. If making a
single-ended measurement, terminate the unused SMA port with a 50-ohm load.
–
3.3. Power LEDs, D3 & D5 should be illuminated when the EVM is connected to power
–
3.4. I
2
C activity can be seen on LED D4 (active with activity).
•
4. Refer to Table 5 for recommended device configurations configurable via the GUI.
Table 5. Recommended Device Configurations
(1)
PARAMETER
156.25 MHz
(2)
100 MHz
161.1328125 MHz
PLL
Reference Doubler
Enabled
N Divider
50
48
51
Fractional Numerator
22500
Fractional Denominator
40000
VCO frequency
5000 MHz
4800 MHz
5162.25 MHz
Output Divider
32
16
32
Loop Filter
Charge Pump Gain
6.4 mA
2.8 mA
Loop Filter Order
2nd Order
3rd Order
C1
5 pF
105 pF
C2
10 nF
C3
0 pF
35 pF
R2
800
Ω
1.1 k
Ω
R3
18
Ω
1.3 k
Ω
Modulator Order
Integer Mode
1st Order
Dithering
Disable
Weak
Charge Pump Bleed
None
8.5 k
Ω
1 ns
(1)
Termination schemes should match device settings. See
(2)
Device default
9
Recommended Test Instruments
For making accurate measurements on ultra-low noise/jitter, high-speed clock signals, the following
instruments are recommended:
•
Source Signal Analyzer: Keysight/Agilent E5052 for phase noise/jitter measurements
•
Oscilloscope: Agilent DSA90000A series (or equivalent) for AC measurements and time-domain jitter
analysis with jitter software package
•
Balun: M/A-COM H-183-4 (30-3000 MHz) 180° coupler, or equivalent
11