SBAS582C – JULY 2014 – REVISED APRIL 2015
8.4.2.8 Reset Program Registers (RST)
The devices support a hardware and software reset (RST) mode in which all program registers are reset to their
default values. The devices can be put into RST mode using a hardware pin, as explained in the
section.
The device program registers can be reset to their default values during any data frame by executing a valid
write operation on the command register with a RST command of 8500h, as shown in
. The device
remains in RST mode if no valid conversion command (AUTO_RST or MAN_Ch_n) is executed and SDI remains
low (refer to the
Continued Operation in the Selected Mode (NO_OP)
section) during the subsequent data
frames. When the device operates in RST mode, the program register settings can be updated (as explained in
the
Program Register Read/Write Operation
section) using 16 SCLK cycles. However, if 32 complete SCLK
cycles are provided, then the device returns invalid data on the SDO line because there is no ongoing conversion
in RST mode. The values of the program register can be read normally during this mode. A valid AUTO_RST or
MAN_CH_n channel selection command must be executed for initiating a conversion on a particular analog
channel using the default program register settings.
Figure 87. Reset Program Registers (RST) Timing Diagram
44
Copyright © 2014–2015, Texas Instruments Incorporated