7 Maintenance
87
This involves the Saveris data logger removed from the system not being
replaced and no measurement data being recorded during the calibration and
adjustment.
1
Press the connect key on the Saveris data logger once to enforce a
final data transmission.
2
Send the Saveris data logger to the calibration laboratory, without
removing it as a component in the startup wizard.
3
When the
Missing data logger connection
system alarm appears,
acknowledge this once.
4
Once the Saveris data logger comes back following calibration, link it
back up with the Saveris base or the Saveris converter.
5
The Saveris data logger automatically acquires its former configuration
data and continues measuring.
7.5
Changing batteries
7.5.1
Changing the Saveris base batteries
The battery in the Saveris base is a wearing part which has to be
replaced after approx. 2 years.
If the battery is faulty, it is not possible to guarantee full operability of
the GSM module. In the event of a power failure, data loss cannot be
ruled out for all components.
As soon as a component's battery is no longer fully functional, it triggers
a
Defective battery
system alarm.
The battery (order no. 0515 5021) should then be replaced immediately
to ensure full functionality and data security.
1
Switch off the Saveris base (with the
Info Base
view selected, press
[ESC]
briefly twice).
2
Disconnect Saveris base from the power supply.