Profile Parameters
The following settings apply to the entire profile. This screen
will appear when
Create Profile
is pressed or when you edit a
profile and
View/Edit Details
is pressed.
Name:
User-entered identifier for the profile. The name follows
the profile when it is saved in a file or moved to another
controller.
Password:
User-set code that must be entered prior to editing a protected profile. A password
may consist of up to ten characters, may include letters, numbers or symbols and is case
sensitive.
Log Data:
Select
Yes
if you want to log data during the profile.
Guaranteed Soak Deviation/Guaranteed Soak Values:
The amount by which the actual
chamber temperature or humidity is allowed to differ from the set point for steps with
Guaranteed Soak Enable set to
On
. In such steps when the actual chamber temperature or
humidity differs from the set point by more than this value, the step timer stops running until the
actual chamber temperature or humidity returns to within the band defined by the set point plus
or minus this value. Each profile can have its own set of Guaranteed Soak values.
Step Parameters
After you create a profile with the procedure above, press the
Options
button in the upper left side. Then press
Create Steps
in
the screen that pops up. You will see one step which is a
Soak
type. Pressing the
Soak
button will allow you to change the step
type. If you are editing an existing profile, pressing the step type
of any step will allow you to change its type. Pressing the
Actions
button will allow you to Edit Step Parameters, Delete Step, or Insert Step.
Each new step will automatically be a copy of the previous step. You can then edit the
parameters of the new step to suit your profile.