Performance Test
45
6.2.5 Test Procedure
Before measuring, check the shielding gasket around the lid. If there is a damaged part,
replace the gasket and wipe the contact area, which is between the lid and the body except
the gasket, with alcohol.
1
Turn the spectrum analyzer and signal generator on.
2
Using an RF cable, connect the RF Connector(N Type) on the rear side of TC-5910DP Shield Box
to the RF output port(N Type) of signal generator.
3
Open the Shield Box lid and connect the 900 MHz Source Dipole Antenna to RF Port(SMA Type)
inside the Shield Box. (Place the antenna at the center of the Shield Box.)
4
Set the output of signal generator to CW900 MHz, 10 dBm.
5
Connect the test antenna to the spectrum analyzer.
Summary of Contents for TC-5910DP
Page 1: ...TC 5910DP Pneumatic Shield Box User Manual R20191023...
Page 9: ...PART 1 1 Getting Started...
Page 13: ...PART 2 5 Hardware Description...
Page 17: ...Dimensions 9 2 2 Dimensions 2 2 1 Outer Dimensions...
Page 18: ...10 Dimensions 2 2 2 Inner Dimensions...
Page 23: ...PART 3 15 Putting into Operation...
Page 49: ...PART 4 41 Maintenance...
Page 57: ...Expendables and Accessories 49 Figure 7 1 TC 5910DP Expendables Location...