34
Error Data List
COUNT?
A query to check the number of lid(door) opened.
Type
Query
Input
COUNT?
Response
8 digit number (ex, 00000029)
5.7.2 Configuration Commands
5.7.2.1 Lid Button (OPEN/CLOSE) Configuration Commands
5.7.2.2 Probe Configuration Commands
The air supply will reset whenever the lid or fixture operation mode is changed, resulting in
opening the lid. If the lid is already open, it will pop open amid closing motion.
5.7.2.1 Lid Button (OPEN/CLOSE) Configuration Commands
DUAL HAND?
DUAL HAND ON
DUAL HAND OFF
DUAL HAND?
A query to check the operating mode of the lid.
Type
Query
Input
DUAL HAND?
Response
{ DUAL HAND ON | DUAL HAND OFF }
●
DUAL HAND ON : Both Lid OPEN/CLOSE Buttons should be pressed to close.
●
DUAL HAND OFF: Only one Lid OPEN/CLOSE Button is enough to close.