Chapter 9 - Performing Medical Procedures
Cleaning the Transducers and Brackets
Terason t3000 / Echo Ultrasound System User Guide
225
Correcting Needle Guide Positioning for Biopsies
When you receive needle guides, they are already configured and tested for angle and
depth. The angle is the number of degrees between the X-axis and the Y-axis (the needle
axis). The depth, shown in millimeters, is the point at which the biopsy needle and guide
lines intersect the vertical center line of the 2D image.
You can make marginal changes to the upper and lower limits for angle and depth on the
Needle Guide Error Correction dialog box. Your changes to these settings are visible in the
needle guidelines, and are saved by the system and used for all biopsies until you change
them again.
You can enter a value within these ranges:
• Angle: –2° to 2°
• Depth: –1 mm to 1 mm
To change the needle guide error correction values for any transducer except the biplanar
transducer, complete these steps:
1.
Click
Tools
>
Needle Guide Error Correction
. The Needle Guide Error Correction
dialog box opens.
Needle Guide Error Correction Dialog Box
You can click the
Apply
button to see the effects of your choices without closing the
dialog box. click the
Default
button to reset the values to the factory-set values.
2.
Next to the Angle Correction field, click the left and right arrows to correct the angle
by the number of degrees.
3.
Next to the Depth Correction field, click the left and right arrows to correct the depth
by the number of millimeters.
4.
Click
OK
to save your entries and close the dialog box.
Cleaning the Transducers and Brackets
The transducer must be cleaned and high-level disinfected between patients. For
instructions, refer to
Processing Terason Transducers Between Uses
To sterilize the brackets, complete the procedure recommended by Civco or Protek.