Preface Before You Begin
Safety Cautions and Warnings
Terason t3000 / Echo Ultrasound System User Guide
16
stated standards with these types of devices connected to the system. Any peripheral
device, such as a network connection, etc. connected to the Terason Ultrasound System
must conform to the IEC standards outlined above (i.e., IEC 60601-1, IEC 60601-1-1 2nd
edition, IEC 60601-1-2, and IEC 60601-2-37).
The Terason Ultrasound System (without peripherals) is suitable for use within the patient
environment, as defined by the following:
The patient environment is defined as shown in the following figure.
Patient Environment
Portable and mobile RF communications equipment can affect medical electrical
equipment such as the Terason Ultrasound System and should not be used in the patient
environment.
Electromagnetic Interference (EMI)
Medical electrical equipment such as the Terason Ultrasound System requires special
precautions regarding electromagnetic compatibility, and must be installed and put into
service according to the tables in
Appendix E, Electromagnetic Compatibility Tables
Warning:
The user should never simultaneously make contact with the
patient and the inside of any equipment where a protective cover
of any kind is removed. This includes the protective covers for
the probe holders containing the ultrasound probe and power
modules located at the patient table.
Caution:
The Terason Ultrasound System should not be used adjacent to or
stacked with other equipment. If adjacent or stacked use is necessary,
the system should be observed to verify normal operation in the
configuration in which it will be used.
Patient Table
1.50 m
Overhead view
Side view
2.50m
1.50m
1.50m