Preface Before You Begin
Labels
Terason t3000 / Echo Ultrasound System User Guide
18
Care
Ultrasound transducers and your Terason Ultrasound System are fragile. Handle with care.
If the transducer appears damaged, or misuse of the electronics envelope causes a
malfunction, please return the equipment to Terason for service or replacement.
Immersion
Do not immerse the transducer connector. If the cable connector is immersed,
do not
plug
the connector into the system. Rinse the connector under running water and dry it
thoroughly. If necessary, contact Terason for service.
HF (High Frequency) Surgical Equipment
Labels
Several labels are attached to the Terason system. The following sections describe what
each label means.
Refer to the laptop user guide for information about its labels and regulatory compliance.
Type BF Equipment Applied Part
Terason Ultrasound System Emits Ultrasound Energy
Indoor Use Only
Warning:
Do not drop the ultrasound transducer! If you accidentally drop
the transducer, check it carefully for damage. Perform a sample
scan to make sure it operates correctly. If not, contact Terason for
service or replacement.
Warning:
The Terason Ultrasound System is not approved for use with HF
surgical equipment. Use of the Terason Ultrasound System in HF
surgical procedures could result in failure of the electrical
isolation, resulting in injury or death.
The Terason Ultrasound System provides protection against electric shock. It is a
Type BF applied part that has a lower leakage current than for non-BF applied
parts.
This device complies with FDA Track 3 guidelines on acoustic output labeling
for the intended uses of the transducers.
This device is intended for indoor use only.