13
14
8. KEYPAD AND DISPLAY
Mode Selection Key
Press this key to select the therapy Mode:
TENS
EMS I, II and III (small, medium and large muscles)
IFT
Microcurrent
When a programme is running, this key also acts a PAUSE button. The PAUSE
symbol
II
will be displayed and the programme timer will stop. Pressing again
resumes the programme and the intensity gradually returns to the set value.
Programme Selection
Press the “Prog” key to select the programme you require (see section 15)
Parameter Menu Selection
Press these keys to select the following parameters one by one:
TENS Preset Mode (P1-P10) :
Treatment Timer (min)
TENS Manual Mode (P11-P12) :
FREQUENCY (Hz)
PULSE WIDTH (μs)
Treatment Timer (min)
EMS Preset Mode (P1-P09) :
Treatment Timer (min)
EMS Manual Mode (P10-P11) :
FREQUENCY (Hz)
PULSE WIDTH (μs)
SYNCHRONOUS/ALTERNATING - SYNCH/ALT in display
WORK TIME (sec)
REST TIME (sec)
RAMP UP/DOWN TIME (sec)
Treatment Timer (min)
ON/OFF
Channel 1
Intensity mA
Lock
Work/Rest/
Ramp Setting
Channel 2
Intensity mA
Lead Fault
Pause
Frequency
Mode
Pulse Width
Programme
Battery State
Warning
Timer
Keys:
ON/OFF Key
(On top of unit)
This key switches the unit on or off.
Press once for 2 sec and the unit is on, the LCD display located at the front of
the unit will light up.
There will be no feeling from either lead at this point as the intensity always
starts at zero. Press this key again and the unit will switch off.
flexistim instructions final.indd 16-17
27/11/2014 12:07