12.2 EMS
Press
MODE
button until EMS I, II, or III selected.
I, II and III are for small, medium, and large muscles and use different Pulse Width settings
P1 appears
Press
PROG
to Select P10 or 11 MANUAL
Press
MENU +
: Hz flashes
Menu steps:
FREQUENCY (Hz)
PULSE WIDTH (μs)
SYNCHRONOUS/ALTERNATING
(SYNCH/ALT in display)
WORK TIME (sec)
REST TIME (sec)
RAMP UP/DOWN TIME (sec)
Treatment Timer (min)
Press
+/-
to adjust the setting
Press
MENU +/-
to save and advance to next Parameter.
A warning triangle will be displayed when the Rest period is less than Rest
Time = Work Time X (Work Hz-16.66)/16.66 seconds. You will not be able to increase the
intensity above zero. When setting the Rest time, you can over-ride this by pressing the
LOCK key , but be aware that using settings outside of this ratio may cause muscle
fatigue. The triangle will continue to flash while you are outside recommended settings.
When using ALTERNATING, ensure that the “Ramp Up Time”, “Work Time”, and “Ramp
Down Time” equals “Rest Time”. The will appear on your screen if this formula is not
followed in EMS manual mode.
If the warning triangle appears at any other time, see “Troubleshooting”.
Adjust intensity above zero to start an EMS programme.
The EMS programmes have three phases: Warm / Train / Cool. You can jump to the next
phase by pressing the SKIP key.
12. ADJUSTING MANUAL PROGRAMMES
Press
PROG
to select one of the manual programmes
Press
MENU+
key to cycle through the available parameters (see list below)
Use the Parameter Adjustment controls to select the required value, then press
MENU+
or
MENU-
to move to the next parameter.
12.1 TENS
Press
MODE
button until TENS is selected
P1 appears
Press
PROG
to Select P11 MANUAL
Press
MENU+
: Hz flashes
Menu steps:
FREQUENCY (Hz)
PULSE WIDTH (μs)
Treatment Timer (min)
Press
+/-
to adjust Parameter
Press
MENU +/-
to save and advance to next Parameter.
Press any intensity button or wait 10 sec to exit setting menu.