11.7 Skip Function
The EMS programmes have three phases:
• Warm
• Train
• Cool
You can jump to the next phase by pressing the SKIP key.
11.6 Adjust the Intensity
Use the left hand keys to adjust the intensity in the left hand lead (Ch1) and the
right hand keys to adjust the intensity in the right hand lead (Ch2).
In IFT MODE both channels are linked and there is only one intensity adjustment.
You can adjust a desired intensity by pressing either of the intensity controls.
For your safety, when intensity reaches 40mA, the warning triangle flashes and
intensity cannot be increased.
Check electrode pad position. If you are certain that the pads are positioned safely - not
through the chest, head or neck - press the SKIP key to override.
The triangle will stop flashing and intensity can be increased to 60mA.
11.4 Set Treatment Timer
The Treatment Timer defaults to Continuous (C) for some TENS programmes and a
safe standard treatment time for others. If you wish to adjust the Treatment Timer proceed
as follows;
To set a Treatment Time, Press either of the Parameter Selection keys:
MENU +/-
The Timer symbol will flash
Use the Parameter Adjustment Controls
+/-
to select your desired treatment
period ranging from 1 to 90 minutes.
Press any intensity key or wait 10 seconds to return to the main screen
The Treatment Timer starts counting as soon as you increase the intensity above zero.
At this point, the display begins to count down from its preset value.
When the preset treatment period is elapsed, the device switches off its output.
11.5 Attach the Electrode Pads
Attach the leads as shown in section 10 and position the electrode pads as shown in
section 16.
23
24
Automatic keypad lock
There is an automatic keypad lock if no button is used for 10 seconds.
Key symbol appears.
Press the Intensity Down button for either channel to unlock.
flexistim instructions final.indd 26-27
27/11/2014 12:07