Check electrode pad position. If you are certain that the pads are positioned
safely, press the SKIP key to override.
The triangle will stop flashing and intensity can be increased to 60mA.
SKIP key
The EMS programmes have three phases: Warm/ Train/ Cool.
You can jump to the next phase by pressing the Skip key.
In IF MODE, use this key to override limit and set intensity at > 40mA.
15
16
IFT Mode
IFT Preset (P1-P3)
Treatment Timer (min)
IFT Manual (P4)
FREQUENCY (Hz) (P1-P3)
Treatment Timer (min)
Microcurrent Mode:
Preset (P1, P2 and P4)
Manual (P3)
WAVEFORM - Const /Square/Ramp
FREQUENCY (Hz)
PULSE WIDTH mS
Treatment Timer (min)
Parameter Adjustment Controls
Press these keys to increase or decrease the value of the parameter which
you have selected with the MENU keys.
Intensity Controls
Press the keys on the left to adjust the intensity in Ch1 and on the right to
adjust the intensity in Ch2
IF MODE:
I.F. signals penetrate deep into the tissue. Positioning pads across the chest
or head could be dangerous - see section 16.3.
For your safety, when intensity reaches 40mA, the warning triangle flashes
and intensity cannot be increased.
Manual Programme Lock
When “Manual” is showing, you can protect the manual settings by pressing
and holding this key for 3 seconds.
The key symbol will flash and you will be unable to change the manual settings.
Press and hold for 3 seconds to unlock programme.
Lead Fault Detector
In TENS and EMS modes, when one or both of the electrodes are not placed firmly
on the skin, or the leads are not properly connected and the intensity is >10, the
output level goes immediately to zero and a lead fault icon flashes.
Automatic keypad lock
There is an automatic keypad lock if no button is used for 10 seconds.
Key symbol appears.
Press the Intensity Down button for either channel to unlock.