System Configurations
74
To reset setting values (MEAS SETUP) and offset values
(OPEN SHORT) to factory default, use the DEFAULT
SET setting. The device can be quickly restored to its
factory settings.
Default settings:
(MEAS SETUP):
FUNC: Cp-D
FREQ: 1kHz
LEVEL: 1V
RANGE: AUTO
TRIG: INT
SPEED: SLOW
SRC RES: 100
BIAS: OFF
AUTO LCZ: OFF
DELAY: 0ms
ALC: OFF
AVG: 1
MON1: OFF
MON2: OFF
(OPEN SHORT):
OPEN TEST: ON
SHORT TEST: ON
SPOT: OFF
Step 1. Press the [Measure] or [Setup] key.
Step 2. Press the [SYSTEM] bottom soft-key.
Step 3. Use the cursor key to select the
[DEFAULT SET] field.
Step 4. Use the soft-keys to set.
Soft-key
Function
ON
To reset to default factory settings
OFF
No reset to factory default settings