T3LCR Series User Manual
35
The T3LCR
OPEN correction capability cancels
errors due to the stray admittance (G, B) in parallel with
the device under test (Refer to Figure 4-3).
The OPEN correction data is taken at all preset
frequency points, independent of the test
frequency you
set. Refer to the tables on page 130 for the trimming
frequency point lists for each model.
Step 1. Press the [Measure] key.
Step 2. Press the [OPEN SHORT] soft-key.
Step 3. Use the cursor key to select the [OPEN] field.
Soft-key
Function
ON
Enables open correction.