The TRIGger subsystem command group is used to enable a measurement or
a sweep measurement, and to set the trigger mode.
The TRIGger:IMMediate command causes the trigger to execute a
measurement or a sweep measurement, regardless of the trigger state.
Command
Syntax
TRIGger[:IMMediate]
Example
SEND> TRIG<NL>
Note
This command can be ONLY used in BUS trigger mode.
The TRIGger:SOURce command sets the trigger mode.
Command
Syntax
TRIGger:SOURce {INT,MAN,EXT,BUS}
Parameter
Where, {INT,MAN,EXT,BUS} is
INT
Internal Trigger Mode
MAN
Manual Trigger Mode
EXT
External Trigger Mode
BUS
BUS Trigger Mode
Example
SEND> TRIG:SOUR BUS<NL>
Query Syntax
TRIGger:SOURce?
Query Response {INT,MAN,EXT,BUS}
Example
SEND> TRIG:SOUR?<NL>
RET>
INT<NL>
The TRIGger:DELAY command sets the trigger delay time.
Command
Syntax
TRIGger:DELAY {<float>,min,max}
TRIGger:DLY {<float>,min,max}
TRIGger
[:IMMediate]
:SOURce
{INT,MAN,EXT,BUS}
:DELAY(DLY)
<float value>,min,max
www.
.com
1.800.561.8187
Summary of Contents for T3LCR1002
Page 1: ...www com information itm com 1 800 561 8187 ...
Page 2: ...www com information itm com 1 800 561 8187 ...
Page 85: ...www com information itm com 1 800 561 8187 ...
Page 96: ...Remote Control 94 www com information itm com 1 800 561 8187 ...
Page 98: ... www com information itm com 1 800 561 8187 ...
Page 102: ...www com information itm com 1 800 561 8187 ...
Page 107: ... value MIN MAX www com information itm com 1 800 561 8187 ...
Page 108: ...www com information itm com 1 800 561 8187 ...
Page 112: ...www com information itm com 1 800 561 8187 ...
Page 121: ... ON OFF or 1 0 ON OFF or 1 0 ON OFF or 1 0 value www com information itm com 1 800 561 8187 ...