Command Reference
122
Query Syntax
CORRection:SPOT:STATe?
Query Response {on,off}
Example
SEND> CORR:SPOT:STATe?<NL>
RET>
on<NL>
The CORRection:SPOT:FREQuency command sets the frequency point for
the specified frequency point correction.
Command
Syntax
CORRection:SPOT:FREQuendy <value>
Parameter
value
NR1,NR2 or NR3:Frequecny value.
A suffix multiplier can be used with this command. But
Example
SEND> CORR:SPOT:FREQ 100<NL>
SEND> CORR:SPOT:FREQ 10k<NL>
Query Syntax
CORRection:SPOT:FREQuency?
Query Response <NR3>
Example
SEND> CORR:SPOT:FREQ?<NL>
RET>
1.03<NL>
This command executes the OPEN correction data measure for the specified
frequency correction.
Command
Syntax
CORRection:SPOT:OPEN
Example
SEND> CORR:SPOT:OPEN<NL>
RET> pass/fail<NL> //Finish or Undone
This command executes the SHORT correction data measure for the
specified frequency correction.
Command
Syntax
CORRection:SPOT:SHORt
Example
SEND> CORR:SPOT:SHOR<NL>
RET> pass/fail<NL> //Finish or Undone
www.
.com
1.800.561.8187
Summary of Contents for T3LCR1002
Page 1: ...www com information itm com 1 800 561 8187 ...
Page 2: ...www com information itm com 1 800 561 8187 ...
Page 85: ...www com information itm com 1 800 561 8187 ...
Page 96: ...Remote Control 94 www com information itm com 1 800 561 8187 ...
Page 98: ... www com information itm com 1 800 561 8187 ...
Page 102: ...www com information itm com 1 800 561 8187 ...
Page 107: ... value MIN MAX www com information itm com 1 800 561 8187 ...
Page 108: ...www com information itm com 1 800 561 8187 ...
Page 112: ...www com information itm com 1 800 561 8187 ...
Page 121: ... ON OFF or 1 0 ON OFF or 1 0 ON OFF or 1 0 value www com information itm com 1 800 561 8187 ...