Measure Key
26
To achieve the best measurement result, it is
recommended to set the impedance range mode to
When using
measurement results may occur if the measured value
exceeds the full scale of the selected range.
Mode
Function
overview
Advantage
Disadvantage
Auto
Range
T3LCR Series
sets the
optimum
impedance
range for the
impedance of
the DUT
need to select
range
The
measurement
time is longer
due to the
ranging time
Hold
Range
Measurement No ranging
time is
required
You need to
select a proper
range
depending on
the value of
the DUT.
Nominal
Range
T3LCR Series
sets the
optimum range
depending on
the nominal
value
need to select
range. No
ranging time is
required
Valid only in
the sorting
mode
The hold range has nine impedance ranges: 10, 30, 100,
300, 1k, 3k, 10k, 30k and 100k.
The impedance range is selected according to the
impedance even if the measurement parameter is
capacitance or inductance.
www.
.com
1.800.561.8187
Summary of Contents for T3LCR1002
Page 1: ...www com information itm com 1 800 561 8187 ...
Page 2: ...www com information itm com 1 800 561 8187 ...
Page 85: ...www com information itm com 1 800 561 8187 ...
Page 96: ...Remote Control 94 www com information itm com 1 800 561 8187 ...
Page 98: ... www com information itm com 1 800 561 8187 ...
Page 102: ...www com information itm com 1 800 561 8187 ...
Page 107: ... value MIN MAX www com information itm com 1 800 561 8187 ...
Page 108: ...www com information itm com 1 800 561 8187 ...
Page 112: ...www com information itm com 1 800 561 8187 ...
Page 121: ... ON OFF or 1 0 ON OFF or 1 0 ON OFF or 1 0 value www com information itm com 1 800 561 8187 ...