The accuracies of L, C, R, |Z| is equal to A
e
, which is
defined by:
A
e
= ±[A×A
r
+(K
a
+K
b
+K
f
)×100+K
L
]×K
c
[%]
A: Basic measurement accuracy
A
r
: Basic accuracy correction factor
K
a
: Impedance factor a
K
b
: Impedance factor b
K
c
: Temperature factor
K
f
: Open/short trimming factor
K
L
:
The method to calculate the accuracy of L and C depends
on whether D
x
The method to calculate the accuracy of R depends on
whether Q
x
When D
x
>
e
, should be
multiplied by
2
1
x
D
.
When Q
x
>
factor, A
e
, should be
multiplied by
2
1
x
Q
.
D
e
=
100
Ae
(when D
x
When D
x
>0.1, D
e
should be multiplied by
(1+D
x
)