Setup Key
56
The T3LCR Series simultaneously measures four
components of the complex impedance (parameters) in a
measurement cycle. These include primary parameter,
secondary parameter and two monitor parameters. Please
refer to the 4.1 [MEAS DISPLAY] Page section on 23 for
details.
T3LCR Series built-in comparator can sort DUTs into a
maximum of 10 bins (BIN1 to BIN9 and OUT), using a
maximum of nine pairs of primary parameter limits and
one pair of secondary parameter limits.
A DUT can be sorted into the auxiliary BIN (AUX)
when the primary parameter is within the limits but the
secondary parameter is not.