Chapter 7
Troubleshooting
75
4. Save the edited file.
The next time WinView is launched, the new name will be displayed on the Hardware
Setup dialog as illustrated in
.
Figure 7-3: Updated Camera Name in
Camera Name
Field
NOTE:
If the Camera Detection Wizard is run again at a later time,
the name will be changed back to the default name (i.e.,
Camera1.)
7.3
Changing the ST-133 Line Voltage and Fuses
The appropriate voltage setting for your country is set at the factory and can be seen on
the power module. If your voltage source changes, you will need to change the voltage
setting and you may need to change the fuse configuration. Refer to
for fuse
rating information.
CAUTION!
!
Use proper fuse values and types for the controller and
detector to be properly protected.
4
4
11
-00
9
7_
0037
Table 7-2: Fuse Ratings
Input Voltage
Fuse
Type
Left
Right
100/120 V
AC
0.75 A
2.50 A
Slo-Blo
220/240 V
AC
0.30 A
1.25 A
Slo-Blo
Summary of Contents for Princeton Instruments PI-MTE
Page 1: ...www princetoninstruments com PI MTE System Manual 4411 0097 Issue 6 October 1 2019...
Page 22: ...22 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 25: ...Chapter 3 Installation 25 PI MTE System Diagram 4411 0097_0014...
Page 26: ...26 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 58: ...58 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 90: ...90 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 94: ...94 PI MTE System Manual Issue 6 Figure B 2 PI MTE Camera 8 01 05 and later 4411 0097_0054...
Page 95: ...Appendix B Outline Drawings 95 Figure B 3 PI MTE Camera 8 01 05 and earlier 4411 0097_0055...
Page 102: ...102 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 112: ...112 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 117: ...This page is intentionally blank...