111
Appendix E:
USB 2.0 Limitations
The following information details the currently known limitations associated with
operation using the USB 2.0 interface.
•
Maximum cable length is 5 meters (16.4 feet);
•
2 MHz is currently the upper digitization rate limit for the ST-133 Controller;
•
Large data sets, and/or long acquisition times, may be subject to data overrun
because of host computer interrupts during data acquisition;
•
USB 2.0 is not supported by the Teledyne Princeton Instruments PC Interface
Library (Easy DLLS);
•
Some WinView 2.5.X features are not fully supported with USB 2.0. Refer to
for complete information.
Table E-1: WinView 2.5.x Feature Support by USB 2.0
Feature
Supported by USB 2.0
Remarks
Demo Port Capability
NO
DIF
NO
Kinetics
YES
WinX 2.5.18.1
Reset Camera to NVRAM Defaults
NO
Temperature Lock Status
YES
WinX 2.5.x doesn’t utilize
hardware lock status
PTG
YES
Virtual Chip
NO
Custom Timing
YES
WinX 2.5.18.1
Custom Chip
YES
WinX 2.5.18.1
Frames per Interrupt
NO
RS170 (Video Output)
NO
Online Exposure
YES
File Information
YES
Not all header info is currently
available in WinX 2.5.x through
PVCAM
Overlapping ROIs
NO
Macro Record
YES
Macros recorded for
non-PVCAM cameras may
have to be re-recorded to
function
TTL I/O
NO
Summary of Contents for Princeton Instruments PI-MTE
Page 1: ...www princetoninstruments com PI MTE System Manual 4411 0097 Issue 6 October 1 2019...
Page 22: ...22 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 25: ...Chapter 3 Installation 25 PI MTE System Diagram 4411 0097_0014...
Page 26: ...26 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 58: ...58 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 90: ...90 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 94: ...94 PI MTE System Manual Issue 6 Figure B 2 PI MTE Camera 8 01 05 and later 4411 0097_0054...
Page 95: ...Appendix B Outline Drawings 95 Figure B 3 PI MTE Camera 8 01 05 and earlier 4411 0097_0055...
Page 102: ...102 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 112: ...112 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 117: ...This page is intentionally blank...