98
PI-MTE System Manual
Issue 6
3. Using two appropriately sized open-end wrenches, place one wrench on the male
fitting and hold it in place while using the second wrench to rotate the female
fitting 90º CW.
NOTE:
For stainless steel or nickel gaskets, rotate 45º CW.
See
Figure C-4: Final Tightening of Fittings
C.1.2
Gasket Replacement Procedure
To ensure proper sealing, gaskets should be replaced before re-connecting previously
connected fittings.
Perform the following procedure to replace the gasket:
1. Place one wrench on the male fitting and hold it in place as you use the other
wrench to rotate the female fitting CCW.
Figure C-5: Initial Loosening of Connection
2. Fully loosen the connection.
Figure C-6: Loosen and Separate Connectors
4
4
11
-0
09
7_
006
1
4
4
11
-00
9
7_
006
2
PI-MTE C
OOLANT
P
IPE
4
4
11-
009
7_
00
5
9
Summary of Contents for Princeton Instruments PI-MTE
Page 1: ...www princetoninstruments com PI MTE System Manual 4411 0097 Issue 6 October 1 2019...
Page 22: ...22 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 25: ...Chapter 3 Installation 25 PI MTE System Diagram 4411 0097_0014...
Page 26: ...26 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 58: ...58 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 90: ...90 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 94: ...94 PI MTE System Manual Issue 6 Figure B 2 PI MTE Camera 8 01 05 and later 4411 0097_0054...
Page 95: ...Appendix B Outline Drawings 95 Figure B 3 PI MTE Camera 8 01 05 and earlier 4411 0097_0055...
Page 102: ...102 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 112: ...112 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 117: ...This page is intentionally blank...