Once you have entered the acquisition parameters, wait for temperature lock to occur
and then wait 30 minutes for the camera temperature to completely stabilize. Then try
taking a few dark charge readings.
NOTES:
1. Do not be concerned about either the DC level of this
background or its shape unless it is very high, i.e., >400
counts. What you see is not noise. It is a fully
subtractable readout pattern. Each CCD has its own
dark charge pattern, unique to that particular device.
Every device has been thoroughly tested to ensure its
compliance with Teledyne Princeton Instruments'
demanding specifications.
2. Do not adjust the offset values to zero or some
low-level data will be missed.
CAUTION!
!
If you observe a sudden change in the baseline signal, TURN
OFF THE SYSTEM IMMEDIATELY and contact the factory
Customer Support Department. Refer to
on page 116 for complete contact information.
Summary of Contents for Princeton Instruments PI-MTE
Page 1: ...www princetoninstruments com PI MTE System Manual 4411 0097 Issue 6 October 1 2019...
Page 22: ...22 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 25: ...Chapter 3 Installation 25 PI MTE System Diagram 4411 0097_0014...
Page 26: ...26 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 58: ...58 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 90: ...90 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 94: ...94 PI MTE System Manual Issue 6 Figure B 2 PI MTE Camera 8 01 05 and later 4411 0097_0054...
Page 95: ...Appendix B Outline Drawings 95 Figure B 3 PI MTE Camera 8 01 05 and earlier 4411 0097_0055...
Page 102: ...102 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 112: ...112 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 117: ...This page is intentionally blank...