31
Chapter 5: IsoPlane Setup
This chapter provides information about the initial setup and configuration of an IsoPlane
spectrograph.
REFERENCES:
For detailed information about the operation of an IsoPlane
, on page 43.
5.1 Unpack and Inspect Equipment
Carefully unpack and examine the IsoPlane and all purchased accessories.
NOTES:
1.
Report any damage immediately to the carrier and to
Princeton Instruments.
2.
Save all packing material.
5.2 Initial Setup Procedure
Perform the following procedure to setup the IsoPlane for the first time:
1.
Position the IsoPlane where it will be used.
2.
Connect the power supply shipped with the IsoPlane to the
POWER
connector located
on the side of the IsoPlane. See
Figure 5-1: External Connector Panel
The power supply pr5 V
DC
for the control logic and +24 V
DC
for the motor
drives.
3.
With the power supply switch in the off position, connect the line cord provided with
the IsoPlane to the power supply and the AC line. The power supply will operate with
line voltages between 100 and 240 V
AC
, 50/60 Hz.
4.
The IsoPlane is designed to be controlled primarily through Princeton Instruments’
LightField
®
or WinSpec data acquisition and control software when operated with a
Princeton Instruments detector. Connect the USB port of the IsoPlane to the USB port
on the control computer using the USB cable provided.
The RS-2332 interface may also be used if a USB port is not available. If using the
RS-232 interface, configure the appropriate port on the host computer as described in
Section D.1, Communication Configuration Settings
44
11-0
143_0
002
POWER
Summary of Contents for Everywhereyoulook IsoPlane 320A
Page 1: ...so lane Manual ...
Page 2: ......
Page 14: ...10 This page is intentionally blank ...
Page 34: ...30 This page is intentionally blank ...
Page 46: ...42 This page is intentionally blank ...
Page 66: ...62 This page is intentionally blank ...
Page 74: ...70 This page is intentionally blank ...
Page 82: ...78 ...
Page 89: ......
Page 90: ......