Chapter 5
IsoPlane Setup
33
b.
Mount the detector to the plate and position the detector so the text on the back of
the detector is readable (i.e., not upside-down.)
NOTE:
The exception to this text-orientation is the PI-MAX2: 1024x256
which should be mounted so that the text is upside-down.
c.
Use a 5/16” open end wrench to tighten each of the three (3) screws.
3.
If the detector is a PyLoN or SPEC-10 LN detector
without shutter, you will need to remove the shroud at
the front of the detector, mount the shroud to the plate,
and then resecure the shroud to the detector. It may be
easier to remove the mounting plate from the IsoPlane
for this operation.
a.
Using a 3/32” ball head hex wrench, remove the
four (4) socket head screws that secure the shroud
to the nose of the detector.
b.
Carefully slide the shroud off.
c.
Remove the mounting plate from the IsoPlane.
d.
Place the front of the shroud against the front of the plate and orient the nose so that
the appropriate hole pattern matches the mounting holes on the plate.
e.
From the inside of the shroud screw three (3) 10-32 x
¼
”, flathead stainless steel
screws through the shroud and into the mounting plate.
f.
Carefully mount the shroud and plate assembly onto the front of the detector and
resecure the shroud with the previously removed four (4) socket head screws.
g.
Mount the plate to the IsoPlane. The detector should be upright.
NOTE:
Refer to the detector’s system manual for specific instructions
about how to mount the array detector to the mounting plate.
These instructions vary with the detector nose design.
5.4 Rotational Alignment
This procedure assumes that you have already mounted the array detector to the exit port of
the spectrograph and that there is a light source at the entrance port. The basic procedure is
provided in this section. Refer to
, on page 43 for specific LightField
and WinSpec instructions for performing a rotational alignment. For other application
software, refer to the specific software’s manual or help file.
Perform the following procedure to rotationally align the detector:
1.
Verify that the computer interface cables for the spectrograph and detector are
connected.
2.
Verify that the power supplies are plugged into an AC source and are connected to the
spectrograph and detector.
3.
Turn on the power to the spectrograph and the detector.
Summary of Contents for Everywhereyoulook IsoPlane 320A
Page 1: ...so lane Manual ...
Page 2: ......
Page 14: ...10 This page is intentionally blank ...
Page 34: ...30 This page is intentionally blank ...
Page 46: ...42 This page is intentionally blank ...
Page 66: ...62 This page is intentionally blank ...
Page 74: ...70 This page is intentionally blank ...
Page 82: ...78 ...
Page 89: ......
Page 90: ......