Automation
Section 4 The Column Selector Module
4-11
in a location that minimizes the possibility of a chemical
spill damaging the power supply.
8. Locate the supplied waste tubing and fitting package (
P
/
N
60-5234-653).
9. Unscrew the attached plastic ferrule in this fitting package
until the plastic ferrule is almost completely detached from
the steel fitting.
10. Attach the steel fitting in this package to the EZ Prep
inject valve waste port. This steel fitting should be hand
tight.
Figure 4-19 Waste tubing fitting attached to the EZ Prep
inject valve
11. Turn the plastic ferrule until it is attached hand tight to
the valve.
12. Turn the steel fitting
1
/
2
turn tighter.
13. Remove the plastic ferrule while leaving the fitting
attached.