Section 4 The Column Selector Module
4-17
along with the diameter is used to estimate a column’s volume
and suggest a default gradient duration. This parameter can also
be overridden to decrease run time.
3. The debubbler automatically operates if one of the names
of the solvents contains the words “water”, “H2O”, “was-
ser”, “eau”, or “agua”. The E
NABLE
D
EBUBBLER
field allows
the user to enable the debubbler without having one of
these words present.
4. To set other method parameters as default for the column
being defined, or to create a new method with custom
parameters, use the D
EFINE
M
ETHODS
button.
5. Once the column definition is set, press the S
AVE
button.
6. To identify the column in each position, select the drop
down menu in upper portion of the screen and choose a
previously defined column. This allows you to select a col-
umn position by selecting the column name in the R
EDI
S
EP
drop down or the AutoSampler Module queue screen.
Figure 4-28 Configuring the Prep HPLC tab