Automation
Section 4 The Column Selector Module
4-6
4.1.2 Establishing Column
Selector Module
Connections
1. Connect the Column Selector Module to the EZ Prep using
the provided USB-to-RS232 cable.
Figure 4-10 Connected USB-to-RS232 cable
2. Remove the stainless steel tubing which connects the EZ
Prep inject valve to the column inlet.
3. Connect one end of the provided stainless steel tubing to
the EZ Prep inject valve.
4. Connect the other end of the provided stainless steel tub-
ing to the inlet port located on the back of the Column
Selector Module.
Figure 4-11 Column Selector Module inlet port
5. Cut a piece of the provided green PEEK tubing which is
long enough to reach from the outlet port of the Column
Selector Module to the outlet port of the EZ Prep.