ADFM® Flow Meter
Section 3 Maintenance and Troubleshooting
3-11
Figure 3-5 Canister controller: opening the electronics lid
2. The interior electronics are attached to the lid. Laying the
controller on its side, very carefully ease the lid a few
inches away from the case (Figure 3-6).
Wiring from the battery compartment is connected to the
electronics.
Do not attempt to pull the electronics all
the way out of the case.
Ensure that the connections between the sensor interface board
and the motherboard are properly seated. There are two connec-
tions: a ribbon cable for VCC and grounding connections, and a
coaxial cable for transmission of the sensor’s analog signal.
Figure 3-6 Canister controller sensor interface board connections
Do not alter the connections made on the motherboard for any
reason other than to re-establish proper connection.
Allen
screws (4)
Ribbon
cable
Coaxial
cable