ADFM® Flow Meter
Section 2 Setup and Installation
2-34
Figure 2-34 Attaching the sensor to a mounting plate
Figure 2-35 Rectangular frame mounting
Where a rectangular frame is used, verify the exact channel
width at different depths and ensure that no obstructions will
hinder insertion and removal of the frame. When designing
mounting methods and accessories, avoid conditions where accu-
mulation of solids on or against the frame is likely. Ensure that
the installation allows access for servicing, clearing of debris, and
removal/replacement of the frame.
Flow
Plate: Part #60-7613-003
FLH M6 X 10mm SST Screws: Part #231-8116-10
Secondary
level sensor
ADFM Pro20
sensor
FLOW